The intuitive design and user-friendly set-up of the Electrosurgical Generator (diathermy) for use in the modern operating room features both Monopolar and bipolar functions to satisfy all surgery demands with safety, flexibility, reliability, and convenience.
The Return electrode contact quality monitoring system (REM) continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected, minimizing potential danger.
Language options include: English, Spanish, Portuguese, Turkish, and more, ensuring global usability.
General Surgery; Gastroenterology; Dermatology; Vascular Surgery; Obstetrics and Gynecology; Heart/Thoracic Surgery; ORL/ENT; Minimally Invasive Surgery (MSI); Cerebral surgery; Neurosurgery, Orthopedics & Plastic Surgery, Trans Urethral Resection (TUR) and more.
Versatile CompatibilityHigh Frequency Leakage: Monopolar: < 150mA; Bipolar < 60mA
Low Frequency Leakage:
Normal Polarity, Intact Chassis Ground: < 10μA
Normal Polarity, Ground open: < 50μA
Reverse Polarity, Ground Open: < 50μA
Sink Current, 140V Applied, All inputs: < 50μA
| Working Mode | Sub-Mode | Max P-P Voltage | Rated Load | Max Output Power | Crest Factor |
|---|---|---|---|---|---|
| Monopolar Cut | Pure | 1900 | 500Ω | 400W | 1.6 |
| Blend | 2300 | 500Ω | 250W | 2.5 | |
| Monopolar Coag | Fulgurate | 2800 | 500Ω | 120W | 2.9 |
| Spray | 7000 | 500Ω | 100W | 6.2 | |
| Bipolar Coag | Standard Coag | 550 | 100Ω | 70W | 1.6 |